1. Type of Configuration: Bench top / freestanding.
2. Type of Bio safety: Vertical direction of airflow.
3. Outer Body: 1.2 mm electro galvanized steel with antimicrobial powder coated finish.
4. Inner Cabinet: Corrosion proof interior design and can be easily cleaned.
5. Work table: High quality rust proof stainless steel of 1.5 mm and type 304.
6. Front manual sliding transparent safety acrylic glass sash to a maximum height for smooth operation. Front window sash opening 10 inches or more.
7. Controller: Microprocessor controlled air velocities in real time with digital display.
8. Intensity of Light in Lux: ≥ 600.
9. Bio Safety Cabinet Has Provision For Fitting Taps: Taps for vacuum, water and non combustible gas.
10. Class of Bio safety cabinet: Class II, Type A2 with approx. 70% recirculation and 30% exhaust.
11. Down flow and exhaust ULPA filter with 99.999% typical efficiency for particle size of 0.1 to 0.2 microns, providing superior product protection over conventional HEPA filters.
12. Blower motors: Quiet in operation and vibration free with noice level <65 Dba. Automatically adjust the airflow speed according to filter load without relying on airflow sensor with brushless DC/ ECM Motor.
13. UV Light: Programmable UV light enables timed sterilization to run to be preset before work commences.
14. Ventilation System: Exhaust and Inflow air volume approx 300-350 CFM.
15. Drain: Valve underneath worktable for collecting sample spillage.
16. Stand: Floor-stand with caster wheels with height of stand of 30 inch+/-1.5 inch.
17. Input power supply requirements: 240 V AC ± 10 %, 50 Hz, single phase.
Technical Data& Ordering Information:-
Model | CLW-153-A | CLW-153-B | CLW-153-C | CLW-153-D | CLW-153-E |
Inner Size (WxDxH) | 2×2×2 Feet | 3×2×2 Feet | 4x2x2 Feet | 6x2x2 Feet | 3x3x3 Feet |
Optional Feature:
1. Stainless steel 316 grades.
2. Intensity of Light in Lux: ≥ 900.
3. Magnehelic Gauge Provided.
4. Hepafilter.
5. Gas Burner.
6. Motorized Sliding Door.
7. Alarm: Audio visual alarm for filter failure.
8. LCD Display for Velocity, Time & Lamp On::Off.
9. UV Hour Meter.
10. UV Germicidal Lamp.
11. DOP HEPA Filter Testing .
12. Digital Pressure Gauge with Alarm.
13. Availability of Display.
14. stabilizer
15. RS 232 Serial Interface Port.